Ji Young Moon (Ewha Womans University School of Medicine) Soo Chang Cho (Ewha Womans University School of Medicine) Hyun Jin Kim (Ewha Womans University School of Medicine) 전루민 (이화여자대학교) 한경은 (이화여자대학교)
저널정보
대한안과학회 Korean Journal of Ophthalmology Korean Journal of Ophthalmology 제36권 제4호
발행연도
2022.8
수록면
326 - 337 (12page)
DOI
10.3341/kjo.2022.0017

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
Purpose: To evaluate the level of agreement between ANTERION (Heidelberg Engineering, Heidelberg, Germany), OA-2000(Tomey, Nagoya, Japan), and IOLMaster 500 (Carl Zeiss AG, Jena, Germany).Methods: Fifty-one eyes of 51 patients were included in the study. Flat keratometry (K) and steep K, vector componentof astigmatism (Jackson cross-cylinder at 0° and 90° [J0] and Jackson cross-cylinder at 45° and 135° [J45]), anterior chamberdepth, and axial length were compared using the three devices. Repeated measures analysis of variance was conducted tocompare the mean values of the biometrics. Pearson correlation test was conducted to analyze the correlations of the measuredvalues, and a Bland-Altman plot was used to assess the agreement between the three devices. The predicted intraocularlens power of each device was compared to the others using the SRK/T, Haigis, Barrett Universal II, and Kane formulas.Results: All K values measured using ANTERION were flatter than those of other instruments. However, good agreement wasobserved for flat K (ANTERION - OA-2000; 95% limits of agreement [LoA], 0.86 diopters [D]) and steep K (ANTERION - OA-2000; 95% LoA, 0.93 D) and OA-2000 - IOLMaster 500 (95% LoA, 0.93 D). J0 and J45 vector components of astigmatism werenot statistically different; however, the agreements were poor between the devices (95% LoA ≥1.97 D). Anterior chamberdepth values of ANTERION and OA-2000 were interchangeable (95% LoA, 0.15 mm). The axial length showed a high agreement(95% LoA ≤0.17 mm) among the three devices. The predicted intraocular lens powers of the three devices were notinterchangeable regardless of formulas (95% LoA ≥1.04 D).Conclusions: Significant differences in ocular biometrics were observed between ANTERION and the other two devices. Thisstudy demonstrated that only axial length showed good agreement among devices.

목차

등록된 정보가 없습니다.

참고문헌 (0)

참고문헌 신청

함께 읽어보면 좋을 논문

논문 유사도에 따라 DBpia 가 추천하는 논문입니다. 함께 보면 좋을 연관 논문을 확인해보세요!

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0