Katharina Hollerieth (1Nephrology Department Klinikum Rechts der Isar of the Technical University of Munich) Bernhard Gaßmann (Meso International GmbH Berlin;) Stefan Wagenpfeil (Institute for Medical Biometry Epidemiology and Medical Informatics Saarland University) Stephan Kemmner (Department of Nephrology Klinikum Rechts der Isar of the Technical University of Munich) Uwe Heemann (Nephrology Department Klinikum Rechts der Isar of the Technical University of Munich) Konrad Friedrich Stock (Nephrology Department Klinikum Rechts der Isar of the Technical University of Munich)
저널정보
대한초음파의학회 ULTRASONOGRAPHY ULTRASONOGRAPHY Vol.37 No.2
발행연도
2018.1
수록면
140 - 148 (9page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
Purpose: This study was conducted to determine the influence of standoff material on acoustic radiation force impulse (ARFI) measurements in an elasticity phantom by using two different probes. Methods: Using ARFI elastography, 10 observers measured the shear wave velocity (SWV, m/sec) in different lesions of an elasticity phantom with a convex 4C1 probe and a linear 9L4 probe. The experimental setup was expanded by the use of an interposed piece of porcine muscle as standoff material. The probe pressure on the phantom was registered. Results: Faulty ARFI measurements occurred more often when quantifying the hardest lesion (74.0 kPa 4.97 m/sec) by the 9L4 probe with the porcine muscle as a standoff material interposed between the probe and the phantom. The success rate for ARFI measurements in these series was 52.4%, compared with 99.5% in the other series. The SWV values measured with the 9L4 probe were significantly higher (3.33±1.39 m/sec vs. 2.60±0.74 m/sec, P<0.001 in the group without muscle) and were closer to the reference value than those measured with the 4C1 probe (0.25±0.23 m/sec vs. 0.85±1.21 m/sec, P<0.001 in the same group). The SWV values measured when using the muscle as a standoff material were lower than those without the muscle (significant for 9L4, P=0.040). The deviation from the reference value and the variance increased significantly with the 9L4 probe if the muscle was in situ (B=0.27, P=0.004 and B=0.32, P<0.001). In our study, the pressure exerted by the operator had no effect on the SWV values. Conclusion: The presence of porcine muscle acting as a standoff material influenced the occurrence of failed measurements as well as the variance and the accuracy of the measured values. The linear high-frequency probe was particularly affected.

목차

등록된 정보가 없습니다.

참고문헌 (40)

참고문헌 신청

함께 읽어보면 좋을 논문

논문 유사도에 따라 DBpia 가 추천하는 논문입니다. 함께 보면 좋을 연관 논문을 확인해보세요!

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0