Hye-Young Lee (Department of Laboratory Medicine Asan Medical Center University of Ulsan College of Medicine Seoul) Sunyoung Ahn (Department of Laboratory Medicine Asan Medical Center University of Ulsan College of Medicine Seoul) Hyun-Ki Kim (Department of Laboratory Medicine Asan Medical Center University of Ulsan College of Medicine Seoul) Woochang Lee (Department of Laboratory Medicine Asan Medical Center University of Ulsan College of Medicine Seoul) Sail Chun (Department of Laboratory Medicine Asan Medical Center University of Ulsan College of Medicine Seoul) Won-Ki Min (Department of Laboratory Medicine Asan Medical Center University of Ulsan College of Medicine Seoul)
저널정보
대한임상검사정도관리협회 Journal of Laboratory Medicine And Quality Assurance Journal of Laboratory Medicine And Quality Assurance Vol.41 No.1
발행연도
2019.1
수록면
47 - 49 (3page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
The objective of this study was to evaluate the analytical performance of the Stat Profile pHOx Ultra Blood Gas Analyzer (Nova Biomedical, USA), a new blood gas/chemistry analyser, including its precision and linearity, comparison studies, and the carry-over effect of commercial reagents and patient specimens. We assessed all the results on the basis of the Clinical and Laboratory Standards Institute guidelines. The following parameters were assessed: pH, partial pressure of carbon dioxide, partial pressure of oxygen, ionized calcium, ionized magnesium (iMg), and lactate concentration The total imprecision had a coefficient of variation of 0.0%–1.8%, and the linear measurement ranges for each parameter were all acceptable. In comparison with the Nova Critical Care Xpress Analyzer (Nova Biomedical, USA), the results indicated a good agreement, except for iMg. All carry-over ranges were between –0.5% and –1.4%. The Stat Profile pHOx Ultra Blood Gas Analyzer showed good analytical performance in terms of precision, linearity, comparison studies, and carry-over effect. The Stat Profile pHOx Ultra Blood Gas Analyzer can provide reliable measurements across a clinically relevant range and has potential use in laboratory tests.

목차

등록된 정보가 없습니다.

참고문헌 (6)

참고문헌 신청

함께 읽어보면 좋을 논문

논문 유사도에 따라 DBpia 가 추천하는 논문입니다. 함께 보면 좋을 연관 논문을 확인해보세요!

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0