Tsuyoshi Majima (Nagoya University Graduate School of Medicine) Yumi Oota (Oita Oka Hospital) Yoshihisa Matsukawa (Nagoya University Graduate School of Medicine) Yasuhito Funahashi (Nagoya University Graduate School of Medicine) Masashi Kato (Nagoya University Graduate School of Medicine) Hiromitsu Mimata (Oita University Faculty of Medicine) Momokazu Gotoh (Nagoya University Graduate School of Medicine)
저널정보
대한비뇨기과학회 Investigative and Clinical Urology Investigative and Clinical Urology Vol.61 No.6
발행연도
2020.1
수록면
613 - 618 (6page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
Purpose: The aim of this study was to investigate whether data obtained from the Lilium α-200 (Lilium Otsuka Co., Ltd., Japan) correlated with conventional frequency-volume chart (FVC) and post-void residual urine volume (PVR) obtained by urethral catheterization. Materials and Methods: This was a prospective multicentre study. Patients hospitalized for the treatment of lower urinary tract symptoms were included. Patients were evaluated with conventional FVC and Lilium α-200 for 2 days. PVR was measured by urethral catherization after urination at the end of the 2 day evaluation period. Results: A total of 42 patients were enrolled in this study. Voided volume and PVR measured by Lilium α-200 were significantly correlated with voided volume obtained from conventional FVC and PVR measured by urethral catheterization, respectively. There was considerable measurement error in voided volume measured by Lilium α-200 (−21.0±102.0 mL). In contrast, the error between PVR measured by the Lilium α-200 and PVR obtained by urethral catheterization was 2.4±52.0 mL. Additionally, high body mass index, but not sex, benign prostate hyperplasia, time zone of measurement (daytime vs. nighttime), and examiners (a urologist versus other healthcare providers) were significantly associated with inaccurate results in voided volume. Conclusions: Voided volume and PVR measured by the Lilium α-200 were correlated with voided volume obtained from conventional FVC and PVR measured by urethral catheterization, although accuracy of the measurements was not high. The Lilium α-200 is a useful device to easily measure approximate bladder volume.

목차

등록된 정보가 없습니다.

참고문헌 (12)

참고문헌 신청

함께 읽어보면 좋을 논문

논문 유사도에 따라 DBpia 가 추천하는 논문입니다. 함께 보면 좋을 연관 논문을 확인해보세요!

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0