Purpose : For practical application of the MR image for stereotactic radiosurgery, the target point achieved by acquisition of MR image in a relatively homogeneous phantom has to agree with the actual isocenter of irradiation in real radiosurgery and the amount of distortion of the MR image should be known. Materials and Methods : A dosimetric film with a random target point was inserted into a radish vertically and horizontally on axis Z and they were fixed with a headring. After image acquisition by stereotactic radiosurgery planning system, we achieved stereotactic coordinate of the target point and examined irradiation using the coordinate acquired as isocenter. After the irradiation, the film in the radish was developed and processed and the degree of coincidence between the target point marked on the film and the center of the radiation distribution. In order to measure the degree of distortion of the MR image in a different way, an acryl phantom was made and punctures were made at intervals of 1 cm and a drop of oil was dropped into it. Then, it was inserted into the radish vertically and horizontally on axis Z to acquire the MR image. Each coordinate was achieved and the estimation of distortion of MR image was made both in vertical and horizontal directions Results : The film from the radio was developed and for the one inserted vertically on axis Z, there was a good coincidence in the discrepancy between the target point marked on the film and the center of the radiation distribution. For the one inserted horizontally, the discrepancy between them was under 0.5 mm. As a result of estimating distortion of MR image using acryl, the discrepancy was under 0.45 mm in the case of the phantom inserted vertically on axis Z, and that of the one inserted horizontally was 1.4 mm. Conclusion : We were able to confirm good coincidence in homogeneous phantom in actual treatment position of radiosurgery using the MR image and the discrepancy measured in the analysis of distortion of the MR image did not exceed the permissible level. Therefore, it was evident the system of the hospital is suitable for radiosurgery using MR image.

목차

등록된 정보가 없습니다.

참고문헌 (0)

참고문헌 신청

함께 읽어보면 좋을 논문

논문 유사도에 따라 DBpia 가 추천하는 논문입니다. 함께 보면 좋을 연관 논문을 확인해보세요!

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0