Kernich Nikolaus (University of Cologne Faculty of Medicine, Cologne, Germany) Heck Vincent J. (Institute of Molecular and Cellular Anatomy (MOCA), Aachen, Germany) Ott Nadine (University of Cologne Faculty of Medicine, Cologne, Germany) Prescher Andreas (University of Cologne Faculty of Medicine, Cologne, Germany) Eysel Peer (Faculty of Medicine and University Hospital Cologne) Vinas-Rios Juan Manuel (University of Cologne Faculty of Medicine, Cologne, Germany)
저널정보
대한척추외과학회 Asian Spine Journal Asian Spine Journal Vol.18 No.4
발행연도
2024.8
수록면
493 - 499 (7page)
DOI
10.31616/asj.2024.0073

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
Study Design: A prospective experimental study.Purpose: This biomechanical in vitro study aimed to examine the extent to which the use of a rod persuader (RP) leads to additional mechanical stress on the screw–rod system and determine its influence on the bony anchoring of primary pedicle screws.Overview of Literature: Degenerative spine diseases and deformities are the most common indications for the stabilization and fusion of spinal segments. The pedicle screw–rod system is considered the gold standard for dorsal stabilization, and an RP is also increasingly being considered to fit the spondylodesis material.Methods: Ten lumbar spines from body donors were examined. Bisegmental dorsal spinal lumbar interbody fusion of the L3–L5 segments was performed using a pedicle screw–rod system (ROCCIA Multi-LIF Cage; Silony Medical, Germany). In group 1, the titanium rod was inserted without tension, whereas in group 2, the rod was attached to the pedicle screws at the L4 and L5 levels, creating a 5-mm gap. To attach the rod, the RP was used to press the rod into the pedicle screw. The rod was left in place for 30 minutes and then removed.Results: The rod reduction technique significantly increased the mechanical load on the overall construct measured by strain gauges (p<0.05) and resulted in outright implant failure with pedicle screw pullout in 88.9%.Conclusions: In cases where the spondylodesis material is not fully attached within the pedicle screw, an RP can be used with extreme caution, particularly in osteoporotic bones, to avoid pedicle screw avulsion and screw anchor failure.

목차

등록된 정보가 없습니다.

참고문헌 (0)

참고문헌 신청

함께 읽어보면 좋을 논문

논문 유사도에 따라 DBpia 가 추천하는 논문입니다. 함께 보면 좋을 연관 논문을 확인해보세요!

최근 본 자료

전체보기

댓글(0)

0