Lee, Chong-Guk (Department of Pediatrics, Inje University Ilsan Paik Hospital, Inje University College of Medicine) Park, Hyang-Mi (Department of Pediatrics, National Medical Center) Shin, Hye-Jung (Department of Pediatrics, National Medical Center) Moon, Jin-Soo (Department of Pediatrics, Inje University Ilsan Paik Hospital, Inje University College of Medicine) Hong, Young-Mi (Department of Pediatrics, Ewha Womans University Mockdong Hospital, Ewha Womans University School of Medicine) Kim, Nam-Soo (Department of Pediatrics, Hanyang University Seoul Hospital, Hanyang University College of Medicine) Ha, Il-Soo (Department of Pediatrics, Seoul National University Children's Hospital, Seoul National University College of Medicine) Chang, Myeong-Jin (Department of Chronic Disease Investigation, Korea Center for Disease and Prevention) Oh, Kyeong-Won (Department of Chronic Disease Investigation, Korea Center for Disease and Prevention)
저널정보
대한소아청소년과학회 Clinical and Experimental Pediatrics Korean journal of pediatrics 제54권 제11호
발행연도
2011.1
수록면
463 - 469 (7page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
Purpose: To validate the Dinamap ProCare 200 blood pressure (BP) monitor against a mercury sphygmomanometer in children 7 to 18 years old in accordance with the 2010 International Protocol of European Society of Hypertension (ESH-IP2) and the British Hypertension Society (BHS) protocol. Methods: Forty-five children were recruited for the study. A validation procedure was performed following the protocol based on the ESH-IP2 and BHS protocols for children and adolescents. Each subject underwent 7 sequential BP measurements alternatively with a mercury sphygmomanometer and the test device by trained nurses. The results were analyzed according to the validation criteria of ESH-IP2. Results: The mean (${\pm}SD$) difference in the absolute BP values between test device and mercury sphygmomanometer readings was $1.85{\pm}1.65$ mmHg for systolic BP (SBP) and $4.41{\pm}3.53$ mmHg for diastolic BP (DBP). These results fulfilled the Association for the Advancement of Medical Instrumentation criterion of a mean${\pm}$SD below $5{\pm}8$ mmHg for both SBP and DBP. The percentages of test device-observer mercury sphygmomanometer BP differences within 5, 10, and 15 mmHg were 96%, 100%, and 100% for SBP, and 69%, 92%, and 100% for DBP, respectively, in the part 1 analysis; both SBP and DBP passed the part 1 criteria. In the part 2 analysis, SBP passed the criteria but DBP failed. Conclusion: Although the Dinamap ProCare 200 BP monitor failed an adapted ESH-IP2, SBP passed. When comparing BP readings measured by oscillometers and mercury sphygmomanometers, one has to consider the differences between them, particularly in DBP, because DBP can be underestimated.

목차

등록된 정보가 없습니다.

참고문헌 (0)

참고문헌 신청

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0