Cho, Eun Hyun (Department of Pediatrics, Samsung Medical Center, Sungkyunkwan University School of Medicine) Song, Jinyoung (Department of Pediatrics, Samsung Medical Center, Sungkyunkwan University School of Medicine) Kang, I-Seok (Department of Pediatrics, Samsung Medical Center, Sungkyunkwan University School of Medicine) Huh, June (Department of Pediatrics, Samsung Medical Center, Sungkyunkwan University School of Medicine) Lee, Sang Yoon (Department of Pediatrics, Sejong General Hospital) Choi, Eun Young (Department of Pediatrics, Seoul National University Bundang Hospital) Kim, Soo Jin (Department of Pediatrics, Konkuk University Medical Center)
저널정보
대한소아청소년과학회 Clinical and Experimental Pediatrics Korean journal of pediatrics 제56권 제9호
발행연도
2013.1
수록면
396 - 400 (5page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
Purpose: The purpose of this study was to share our experience of transcatheter closure of small patent ductus arteriosus (PDA) by using an Amplatzer vascular plug (AVP). Methods: We reviewed the medical records of 20 patients who underwent transcatheter closure at Samsung Medical Center and Sejong General Hospital from January 2008 to August 2012. The size and shape of the PDAs were evaluated by performing angiograms, and the PDA size and the AVP devices size were compared. Results: The mean age of the patients was $54.9{\pm}45.7$ months old. The PDAs were of type C (n=5), type D (n=12), and type E (n=3). The mean pulmonary end diameter of the PDA was $1.7{\pm}0.6$ mm, and the aortic end diameter was $3.6{\pm}1.4$ mm. The mean length was $7.3{\pm}1.8$ mm. We used 3 types of AVP devices: AVP I (n=5), AVP II (n=7), and AVP IV (n=8). The ratio of AVP size to the pulmonary end diameter was $3.37{\pm}1.64$, and AVP size/aortic end ratio was $1.72{\pm}0.97$. The aortic end diameter was significantly larger in those cases repaired with AVP II than in the others (P=0.002). The AVP size did not significantly correlate with the PDA size, but did correlate with smaller ratio of AVP size to aortic end diameter ($1.10{\pm}0.31$, P=0.032). Conclusion: Transcatheter closure of small PDA with AVP devices yielded satisfactory outcome. AVP II was equally effective with smaller size of device, compared to others.

목차

등록된 정보가 없습니다.

참고문헌 (0)

참고문헌 신청

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0