Kim, Hyuncheol (Environmental Radioactivity Assessment Team, Korea Atomic Energy Research Institute) Jung, Yoonhee (Environmental Radioactivity Assessment Team, Korea Atomic Energy Research Institute) Lee, Wanno (Environmental Radioactivity Assessment Team, Korea Atomic Energy Research Institute) Choi, Guen-Sik (Environmental Radioactivity Assessment Team, Korea Atomic Energy Research Institute) Chung, Kun Ho (Environmental Radioactivity Assessment Team, Korea Atomic Energy Research Institute) Kang, Mun Ja (Environmental Radioactivity Assessment Team, Korea Atomic Energy Research Institute)
저널정보
대한방사선방어학회 방사선방어학회지 방사선방어학회지 제41권 제4호
발행연도
2016.1
수록면
395 - 401 (7page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
Background: Liquid scintillation counters (LSCs) are commonly used as an analytical method for detecting $^{222}Rn$ in groundwater because they involve a simple sample pretreatment and allow high throughout with an autosampler. The Quantulus 1220 is the best-selling LSC in Korea, but its production was stopped. Recently, a new type of LSC, the 300SL, was introduced. In this study, the 300SL was compared with the Quantulus 1220 in order to evaluate the ability of each apparatus to detect $^{222}Rn$ in groundwater. Materials and Methods: The Quantulus 1220 and 300SL were used to detect the presence of $^{222}Rn$. Radon gas was extracted from a groundwater sample using a water-immiscible cocktail in a LSC vial. The optimal analytical conditions for each LSC were determined using a $^{222}Rn$ calibration source prepared with a $^{226}Ra$ source. Results and Discussion: The optimal pulse shape analysis level for alpha and beta separation was 80 for the Quantulus 1220, and the corresponding pulse length index was 12 in the 300SL. The counting efficiency of the Quantulus 1220 for alpha emissions was similar to that of the 300SL, but the background count rate of the Quantulus 1220 was 10 times lower than that of the 300SL. The minimum detectable activity of the Quantulus 1220 was $0.08Bq{\cdot}L^{-1}$, while that of the 300SL was $0.20Bq{\cdot}L^{-1}$. The analytical results regarding $^{222}Rn$ in groundwater were less than 10% different between these LSCs. Conclusion: The 300SL is an LSC that is comparable to the Quantulus 1220 for detecting $^{222}Rn$ in groundwater. Both LSCs can be applied to determine the levels of $^{222}Rn$ in groundwater under the management of the Ministry of Environment.

목차

등록된 정보가 없습니다.

참고문헌 (0)

참고문헌 신청

함께 읽어보면 좋을 논문

논문 유사도에 따라 DBpia 가 추천하는 논문입니다. 함께 보면 좋을 연관 논문을 확인해보세요!

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0