The present work used the efficiency transfer method used to calculate the full energy peak efficiency (FEPE) curves ofthe (2"*2" & 3"*3") NaI (Tl) detectors based on the effective solid angle subtended between the source and the detector. Thestudy covered the effect of the self attenuation coefficient of the source matrix (with a radius greater than the detector's radius)on the detector efficiency. 152 An Eu aqueous radioactive source covering the energy range from 121.78 keV up to 1408.01keV was used. In this study an empirical formula was deduced to calculate the difference between the measured and thecalculated efficiencies [without self attenuation] at low and high energy regions. A proper balance between the measured andcalculated efficiencies [with self attenuation] was achieved with discrepancies less than 3%, while reaching 39% forcalculating values [without self attenuation] due to working with large sources, or for low photon energies.