Oligotrophic microbial fuel cells (MFCs) were testedfor the continuous monitoring of low biochemical oxygendemand (BOD) by using artificial wastewater, containingglucose and glutamate, as check solution. Ten times dilutedtrace mineral solution was used to minimize the backgroundcurrent level, which is generated from the oxidation ofnitrilotriacetate used as a chelating agent. The feeding rate of0.53 ml/min could increase the sensitivity from 0.16 to0.43 μA/(mg BOD/l) at 0.15 ml/min. The dynamic linear rangeof the calibration curve was between 2.0 and 10.0 mg BOD/l,and the response time to the change of 2 mg BOD/l was about60 min. The current signal from an oligotroph-type MFCsincreased with the increase in salts concentration, and the salteffect could be eliminated by 50 mM phosphate buffer.